Introduction TO RX50M Series Metallurgical Microscope
Adhering to the philosophy of "continuous exploration &selftranscending" for SUNNY brand, new RX50M metallurgical microscope forresearch with a number of pioneering design in appearance and functions, isborn to provide a perfect detection solution and develop a new pattern ofindustrial field.
RX50M viewing head is designed of multiple options for splitting ratio.with wide beam imaging system, 26.5mm super wide filed of view is available.
Trinocular head with erect image, splitting ratio Binocular:Trinocular=100:0 or 0:100. The moving direction of samples is as the same as observed.
Trinocular head with inverted image, splitting ratio Binocular:Trinocular=100:0 or 20:80 or 0:100. Except for concentrating 100% light toeyepiece tube or camera tube, there is another option with 20% light toeyepiece tube and 80% to camera tube, so that eyepiece observation and imageoutput can be available at the same time.
Polarization system The polarizer and analyzer in polarization systemcontribute to eliminate the stray light in semiconductor and PCB detection,image with clear detail is achievable. There are fixed analyzer and rotatableanalyzer for option. The sample can be observed at different polarizing angleswith 360° rotatable analyzer. Besides, this polarization system can be upgradedto Nomarski differential interference contrast system after installing a newdeveloped DIC attachment.
Nomarski differential interference contrast system The weeny asperities onthe surface which can not be found in bright field, is able to be detected byusing U-DICR attachment to create high contrast background. It is widely usedfor testing the conductive particle of LCD, surface scratch of precision disk.
Linkage between the neutral density filter and the switch for BF & DFThe lever in front of illuminator is used to switch between bright field anddark field, and it is in tandem with a neutral density filter (ND50). When youswitch from DF to BF, the built-in ND50 filter takes the role to decay thelight intensity. More scientific and more comfortable.
Technical Specifications
Optical system | Infinity color corrected optical system |
Viewing head | Erect image, 30° inclined gemel trinocular head, interpupillary distance: 50mm~76mm; splitting ratio R: T=100:0 or 0:100 |
Inverted image, 30° inclined gemel trinocular head, interpupillary distance: 50mm~76mm; splitting ratio R:T=100:0 or 20:80 or 0:100 | |
Eyepiece | High eyepoint wide field plan eyepiece PL10X25mm, diopter adjustable |
High eyepoint wide field plan eyepiece PL10X25mm, with reticle, diopter adjustable | |
High eyepoint wide field plan eyepiece PL10X26.5mm, diopter adjustable | |
High eyepoint wide field plan eyepiece PL10X26.5mm, with reticle, diopter adjustable | |
Objective | BD semi-apochromatic metallurgical objectives (5X/10X/20X/50X/100X) |
Semi-apochromatic metallurgical objectives (5X/10X/20X/50X/100X) | |
Nosepiece (with DIC slot) | BD quintuple nosepiece, BD sextuple nosepiece |
Sextuple nosepiece, septuple nosepiece | |
Frame | Reflected/Transmitted body, low-position coaxial coarse and fine adjustment, coarse adjustment distance: 25mm; fine precision: 0.001mm. With coarse adjustment stop and tightness adjustment. |
Built-in 100-240V wide voltage transformer, double way power output; intensity adjustable by digital set and reset; switch for reflection and transmission; built-in transmitted filters LBD/ND6/ND25). | |
Reflected body, coaxial coarse and fine adjustment, coarse adjustment distance: 25mm; fine precision: 0.001mm. With coarse adjustment stop and tightness adjustment. | |
Stage | 4 inch three layers mechanical stage with glass plate, moving range: 102mm(Y)*105mm(X) |
Condenser | Swing-out type achromatic condenser (N.A.0.9) |
Reflected illuminator | BD reflected illuminator with iris field diaphragm and aperture diaphragm, central adjustable. With filter slot and polarizing slot. With switch for bright and dark field |
lamp house | 12V/100W halogen lamp house, center pre-set |
Other accessories | Camera adapter: 0.5X focusing C-mount |
Fixed polarizer, fixed analyzer, 360° rotatable analyzer | |
DIC attachment | |
Interference filters for reflected light: | |
High precision micrometer, scale value 0.01mm |